ADC Spectral Testing with SignalAmplitude Drift and Simultaneous Non-coherent Sampling

  • Authors:
    Yuming Zhuang (Iowa St Univ.), Degang Chen (Iowa St Univ.)
    Publication ID:
    P090858
    Publication Type:
    Paper
    Received Date:
    8-May-2017
    Last Edit Date:
    8-May-2017
    Research:
    1836.127 (Iowa State University)

Abstract

Spectral testing is a vital category in modern measurement industry. The challenge is to cost-effectively test the ADCs accurately. Among many test requirements, maintaining input test signal stability and achieving coherent sampling are one of the bottlenecks for accurate testing of the ADC. If these stringent test requirements are not met, accurate spectral test results of the ADC cannot be obtained. This paper presents a method that simultaneously relaxes the requirements of precise control over source amplitude, and of the need for coherent sampling. By dividing the output data into segments, accurate estimation of the drift fundamental in each segment can be obtained, by removing the estimated drift fundamental and replacing with a coherent, non-drift fundamental in time domain, accurate spectral results can be achieved. Extensive simulation results have validated the accuracy and robustness of the proposed method. Furthermore, measurement results have verified the proposed method by accurately obtain spectral results of the ADC under test.

4819 Emperor Blvd, Suite 300 Durham, NC 27703 Voice: (919) 941-9400 Fax: (919) 941-9450